Main Technical Parameters And Performance Indicators



ParametersIndicator
Test range

CHOL:2.59-12.93mmol/L(100-500mg/dL)

HDL:0.39-2.59mmol/L(15-100mg/dL)

TRIG:0.51-7.34mmol/L(45-650mg/dL)

Accuracy (relative deviation %)

CHOL:2.59-12.93mmol/L(100-500mg/dL) (≤15%)

HDL:0.39-2.59mmol/L(15-100mg/dL) (≤15%)

TRIG:0.51-7.34mmol/L(45-650mg/dL) (≤15%)

Repeatability(CV%)

CHOL:5-7mmol/L(193-271mg/dL) (≤10%)

HDL:0.8-1.5mmol/L(31-58mg/dL) (≤15%)

TRIG:1.5-2.5mmol/L(133-221mg/dL) (≤10%)

Linear(r)

CHOL:2.59-12.93mmol/L(100-500mg/dL) (≥0.975)

HDL:0.39-2.59mmol/L(15-100mg/dL) (≥0.975)

TRIG:0.51-7.34mmol/L(45-650mg/dL) (≥0.975)

Sample typeWhole blood (venous blood or capillary blood)
Sample addition quantity35uL
Test time≤120s
Power supply4*1.5V AAA battery
Unit of measurementmmol/L,mg/dL (mmol/L by default)
Storage200 sets of data with the test date and time
Power-saving modeThe device will be turned off automatically if there is no operation within 3 minutes
Device size140mm×82mm×25mm
weightAbout 140g (excluding batteries)
Operating environment10-30℃;humidity:≤80%
Storage environment0.50℃;humidity:≤90%
USB portFor a developer to debug




Technical specification



Test Type

CHOL,HDL,TRIG,CHOL/HDL,LDL

Test Sample

Whole blood(venous or capillary blood),applicable to whole blood contained heparin anticoagulant

Blood Collection

Blood collection needle

Sample Volume

35μL

Detection Range

CHOL: 2.59-12.93mmol/L(100-500mg/dL)

HDL: 0.39-2.59mmol/L (15-100mg/dL)

TRIG: 0.51-7.34mmol/L (45-650mg/dL)

Test Period

≤120 s

Results Storage

200 lasted results

Battery

4 AA/1.5V batteries, and battery life almost 1000 times

Accessories

Quality control test card,Disposable blood sampling needle(Optional), Disposable micro blood collection device(Optional)

Dimensions

140mm x 82mm x 25mm (length x width x thickness)

Weight

About 140g (Without batteries)

Test strip(LP-S3-01)

purchase it separately

LP-S3-01 Total Cholesterol(1/10/ 20/25/ 50 per box)

Triglycerides(1/10/ 20/25/ 50 per box)

High Density Lipoprotein Cholesterol Test Card(1/10/ 20/25/ 50 per box)



Diagnostic Kits

Product Name

Packing Specification


Lipid Profile Test Strip

1 test/bag, 10 tests / box, 20 tests / box, 25 tests

/ box, 50 tests / box




Thank You for Your Attention on Lepu Medical!

Email us with any questions or inquiries or use our contact data. We would be happy to answer your questions.

Submit
Submit